Smiths Medical Issues Urgent Medical Device Correction Letter Notifying Customers of a Potential Issue with Level 1® H-2 Pressure Chambers used with Level 1 Fast Fluid Flow Fluid Warmers


MINNEAPOLIS, Aug. 3, 2022 /PRNewswire/ -- Smiths Medical issued an Urgent Medical Device Correction Letter to notify affected customers of a potential issue with specific Level 1® H-2 Pressure Chambers used with the Level 1 Fast Flow Fluid Warmers.

In 2015, Smiths Medical implemented a design change to widen the hinge/latch assembly on the Level 1 H-2 Pressure Chambers used with the Level 1 Fast Flow Fluid Warmers (Models H-1025 or H-1200) or added it to the H-1000 model. The company has become aware that Level 1 H-2 Pressure Chambers with the wider hinge/latch assembly can potentially impact the pressure exerted onto the IV fluid bag while contained within the pressure chamber, which may result in decreased flow rate, stopped flow, or residual fluid left within the IV bag.

Affected product was distributed in the United States between December 19, 2016, and March 10, 2022.  However, all Level 1 H-2 Pressure Chamber devices have the potential to be affected by this issue, because some devices may have received a hinge/latch replacement during that timeframe. Level 1 H-2 Pressure chambers with the wide hinge/latch assembly are more susceptible to this issue in the following scenarios:

  1. kinked tubing on the disposable administration sets
  2. use of the lowest flow rate disposables (DI-50, D-70, or DI-70) when delivering viscous fluids such as chilled blood from 300 mL or smaller IV bags

Decreased flow rate, stopped flow, or residual fluid left within the IV bag could result in under-delivery or delay of therapy, leading to potential inadvertent hypothermia, hypovolemia, and/or hypotension which may lead to serious injury and death.

Smiths Medical has sent all affected customers and distributors a letter outlining the risk and providing specific steps to determine whether their devices are affected or not and what steps to follow if they have affected product in service. A copy of that letter can be found here

For further inquiries, please contact Smith Medical using the information provided below.

Smiths Medical Contact

Contact Information

Areas of Support

Global Complaint Management

[email protected]

1-(866)-216-8806

To report adverse events or product complaints

Device Correction Inquiries

1-(800)-241-4002, option 4, then prompt 1

For any questions regarding this action

Technical Assistance

1-(800)-258-5361, option 2, then prompt 6

Additional information or technical assistance

The U.S. Food and Drug Administration (FDA) has been notified of this action.

Adverse reactions or quality problems experienced with this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax.

Media Contact:
Tom McCall
949-366-4368
[email protected]

SOURCE Smiths Medical


These press releases may also interest you

at 06:03
The IAMStm brand, proudly part of the Mars family of brands, announced today the launch of the IAMStm Puppy Hub and Kitten Hub, two online tools featuring expert knowledge and advice for new and soon-to-be pet parents. Each Hub will bring...

at 06:00
Today, TELUS Health announced a collaboration with Nova Scotia Health (NSH) to enable residents of Nova Scotia to access their primary care information through the YourHealthNS app. This health data interoperability initiative marks the first...

at 05:30
The global Digital Healthcare Market, valued at $220.10 billion in 2024, is on a trajectory of rapid expansion, with projections indicating it will soar to $836.10 billion by 2031, growing at a compound annual growth rate (CAGR) of 21% from 2024 to...

at 05:09
Medical imaging IT and cybersecurity company Sectra (STO: SECT B) is publishing its Annual Report and Sustainability Report for the 2023/2024 fiscal year today. This report also includes the Corporate Governance Report for the same period. The...

at 03:20
Calliditas Therapeutics AB (STO: CALTX) ("Calliditas") today announces that its partner Viatris Pharmaceutical Japan G.K. ("Viatris") has initiated a phase III clinical trial in Japan with Nefecon, named VR-205 in the Japanese market, in Japanese...

at 03:00
Global biotechnology leader CSL Behring today announced that two hemophilia B patients were treated with the gene therapy HEMGENIX® (etranacogene dezaparvovec) at Hemophilia Treatment Centers in France. This milestone achievement makes HEMGENIX® the...



News published on and distributed by: